Publication: A pipeline for adaptive filtering and transformation of noisy left-arm ECG to its surrogate chest signal
Loading...
Date
2020-05
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
MDPI AG
Abstract
The performance of a low-power single-lead armband in generating electrocardiogram (ECG) signals from the chest and left arm was validated against a BIOPAC MP160 benchtop system in real-time. The filtering performance of three adaptive filtering algorithms, namely least mean squares (LMS), recursive least squares (RLS), and extended kernel RLS (EKRLS) in removing white (W), power line interference (PLI), electrode movement (EM), muscle artifact (MA), and baseline wandering (BLW) noises from the chest and left-arm ECG was evaluated with respect to the mean squared error (MSE). Filter parameters of the used algorithms were adjusted to ensure optimal filtering performance. LMS was found to be the most effective adaptive filtering algorithm in removing all noises with minimum MSE. However, for removing PLI with a maximal signal-to-noise ratio (SNR), RLS showed lower MSE values than LMS when the step size was set to 1 × 10−5. We proposed a transformation framework to convert the denoised left-arm and chest ECG signals to their low-MSE and high-SNR surrogate chest signals. With wide applications in wearable technologies, the proposed pipeline was found to be capable of establishing a baseline for comparing left-arm signals with original chest signals, getting one step closer to making use of the left-arm ECG in clinical cardiac evaluations.
Description
Keywords
Adaptive Filtering, ECG Transformation, Filter Optimization, Left-arm ECG, Single-lead ECG, Wearable ECE Armband